Cool-125 Circular Coil
Product Number: MV9016E0511

- Diameter: 121mm
- Pulses before warmup: >20.000
- Trigger button
- Cooled dynamically with external coil cooler unit
- Compatible with R30 & X100

 

Electrical and magnetic properties similar to the MCF-125. Designed for demanding stimulation protocols requiring a very high number of stimuli. Equipped with trigger button to support clinical operation. Built in EPV timer and counter (5 years or 18 million EPV, whichever comes first).

 

Mechanical Properties:

  • Weight of transducer head 2.5 kg
  • Cable length 1.3 m
  • Dimensions of transducer head ø140 x 45 mm

 

Coil Winding Data:

  • Inner diameter 15 mm
  • Outer diameter 121 mm
  • Winding height 12 mm
  • Number of windings 15

 

Magnetic and Electrical Properties:

  • Max initial dB/dt 34 kT/s near the coil surface.
  • Active pulse width 280µs (Biphasic)

 

Performance:

  • Number of stimulations, before warm-up at ambient temperature 20°C, output 75% of maximum at 2pps.
  • >20.000 pulses
  • Number of stimulations, before warm-up at ambient temperature 20°C with protocol: 60 trains @ 50 pulses/train @ 10pps @ Inter Train Interval: 25s @ Output=75%.
  • >10.000 pulses